+90 232 532 08 32
info@novatech-control.com

1017 VoltageCurrentResistance Calibrator

Anasayfa / Ürünlerimiz / 1017 VoltageCurrentResistance Calibrator

1017 VoltageCurrentResistance Calibrator

Product Description:

A high performance portable multifunction calibrator with voltage, current, and resistance ranges. The 1017 combines precision with simple operation, making it suitable for use in the workshop or field. Constructed in a compact and durable plastic case with a tilt stand/carry handle it takes up minimal bench space and is easily transportable.

 

Five DC voltage ranges from 10mV to 100V full scale are available, each with a 6-digit (1ppm) resolution. The DC current range is 100mA full scale with a 100nA (1ppm) resolution. Resistance from 0.01Ω to 10kΩ is available 0.01Ω steps.

 

The 1017 can be powered from mains supply or by the internal rechargeable battery pack. Battery operation enables good performance where earth loop and noise pick-up occurs. When the calibrator is plugged into the mains supply the internal batteries will automatically start to recharge. If unplugged from the mains during operation the internal batteries will continue to power the instrument. Full charge allows 12 hours typical use. The battery condition monitored by a meter on the front panel.

 

Stability vs Temperature and Time

Outstanding performance is achieved by the use of special computer selected reference diodes and the latest in resistor technology. The special low-thermal emf terminals reduce errors when working with microvolt signals.

 

Digital Deviation Control

Allows the output to be increased/decreased in % terms from 0 to +/-0.999%. This provides a direct read-out of error and simplifies the recording results for calibration certificates. It enables the user to immediately see if the unit under test is within specification.

 

Applications

The 1017 is suitable for calibrating and simulating a wide range of instruments including thermocouples, transducers, 4-20mA and 0-10V transmitters, and platinum resistance thermometers.

 

As part of the Time Electronics multifunction calibrators range the 1017 is designed for accuracy and reliability, whilst maintaining ease of use.

It is a versatile calibration instrument that provides a high performance solution to voltage, current and resistance applications.

 

 

 

Technical Specification:

Voltage

Ranges & Accuracy        0 to 9.99999mV in 10nV steps / ± 0.02% of setting ± 0.005% of range

0 to 99.9999mV in 100nV steps / ± 0.01% of setting ± 0.004% of range

0 to 999.999mV in 1μV steps / ± 0.005% of setting ± 0.002% of range

0 to 9.99999V in 10μV steps / ± 0.005% of setting ± 0.002% of range

0 to 99.9999V in 100μV steps / ± 0.01% of setting ± 0.004% of range

Output Resistance         10mV and 100mV: 10Ω. 1V and 10V: <150mΩ. 100V: <1Ω.

Drive Current Max         10 and 100mV: as 10Ω output resistance. 1V and 10V: 150mA; 100V: 10mA.

Current

Range & Accuracy          0 to 99.9999mA in 0.1μA steps / ± 0.02% of setting ± 0.004% of range

Drive Voltage Max         10V

Resistance

Range & Accuracy          0 to 9.99999kΩ in 0.01Ω steps / ± 0.05% of setting ± 0.003% of range

Power Rating     0.25W per resistor

Residual resistance        Less than 200mΩ

Deviation Control (V&I) 0% to 0.999% in 0.001% steps. Deviation accuracy: V & I output, 0.5%.

Temperature Coefficient            <5ppm per °C

Long Term Stability        <5ppm/day, <15ppm/90day, <25ppm/year

Short Term Stability – Noise       10mV range: <0.2μV/sec, <0.3μV/10sec, <0.4μV/min

100mV range: <0.2μV/sec, <0.4μV/10sec, <0.6μV/min

1V range: <0.2μV/sec, <0.5μV/10sec, <1.5μV/min

10V range: <1.0μV/sec, <2.0μV/10sec, < 8.0μV/min

100V range: <40μV/sec, <100μV/10sec, <500μV/min

100mA range: <0.2μA/sec, <0.4μA/10sec, <1.0μA/min

Warm-up/Settling Time Warm-up: <10 mins to full accuracy. Settling: < 0.5 secs, 100V range 5 secs.

Output Connections      The output is via low thermal emf terminals (0.2μV/°C). A mains earth terminal is provided for screening purposes. Output polarity can be selected by a switch on the front panel.

Power Supply    The 1017 can be powered continuously from a 230V 50/60 Hz (110V to order) mains supply, or from the internal rechargeable NiCad battery pack.

Battery Level Indicator  A front panel display provides a continuous indication of the battery state.

The above accuracies are independent of thermal emfs which can be 2μV or more depending on the type of leads and connections used.

 

Operating Temperature              0 to 50°C (32 to 120°F). 15 to 25°C for optimum performance.

Operating Humidity       10 to 90% non-condensing 25°C (77°F)

Dimensions       W250x H119 x D314mm

Weight 2.4kg

Optional Extras Calibration Certificates – traceable to NPL and UKAS

Country of Origin           UK