Anasayfa / Ürünlerimiz / ISOTECH EUROPA – VENUS – CALISTO 4000 SERIES
These models will calibrate temperature probes from -45°C to 250°C with unrivalled flexibility. As a traditional Dry Block, several thermometers can be quickly calibrated.
Accessories are available to convert to a stirred liquid bath, for surface sensor calibration, to calibrate infrared thermometers and even to use as an ITS-90 Fixed Point System with calibration uncertainties as small as 0.0005°C. With excellent stability and distributed heating cooling zones for good uniformity these calibrators offer proven thermal performance.
These award winning calibrators are easy to use and are available in three versions – the Basic, the Site and the ADVANCED. The Basic has a digital display of set and nominal temperature, the Site additionally includes an inbuilt independent temperature indicator for a reference probe. The ADVANCED controller has inputs for reference and test thermometers with a further range of sophisticated features including automatic temperature cycling, secure data logging and full colour high resolution display.
All models include I-Cal Easy LOG software and the ADVANCED models additionally include software to manage logged data and configure the unit, more details.
These models meet the calibration capacity requirements of EURAMET/cg-13/v.01, “EA Guidelines on the Calibration of Temperature Block Calibrators”, formerly EA10/13. accuracy and best performance.
PARAMETER |
MODEL |
||
---|---|---|---|
Europa 4520 |
Venus 4951 |
Calisto 4953 |
|
Temperature Range |
-45°C to 140°C1 |
-35°C to 140°C2 |
30°C to 250°C3 |
ADVANCED Range |
|||
Stability: Dry Block / Liquid Bath |
±0.01°C |
±0.01°C |
±0.02°C |
Display Resolution |
0.001°C over whole range |
0.001°C over whole range |
0.01°C over whole range |
Input Channel Accuracy: Thermocouple |
E, J, K, N ±0.2°C @ 660°C R: ±0.6°C S: ±0.7°C @ 660°C T: ±0.2°C @ 150°C |
“ |
“ |
CJC Accuracy |
±0.35°C |
“ |
“ |
Input Channel Accuracy: RTD |
±0.05°C ±0.005% RDG |
“ |
“ |
BASIC / SITE Range |
|||
Stability |
±0.03°C |
±0.03°C |
±0.03°C |
Display Resolution |
0.01°C from -19.99 to 99.99°C then 0.1°C: 0.01°C Over PC Interface |
||
COMMON Specifications |
|||
Stability |
Blackbody ±0.03°C Surface Sensor ±0.5°C ITS-90 Cells ±0.0005°C |
“ |
“ |
Unformity – Between Wells Dry Block Mode (Radial) |
<0.008°C |
<0.008°C |
<0.02°C at 250°C |
Uniformity – Lower 40mm (Axial) Dry Block Mode |
<0.040°C |
<0.040°C |
<0.25°C |
Uniformity – Radial Bath Mode |
<0.02°C |
<0.02°C |
<0.011°C at 250°C |
Uniformity – Lower 40mm (Axial) as Liquid Bath |
<0.026°C |
<0.026°C |
<0.02°C at 250°C |
Heating Time |
-30°C to 140°C: 15 mins |
-30°C to 140°C: 15 mins |
25°C to 250°C: 15 mins |
Cooling Time |
140°C to 0°C: 15 mins |
140°C to 0°C: 15 mins |
250°C to 30°C: 25 mins |
Calibration Volume |
35 x 160mm |
“ |
“ |
Standard Insert |
6 pockets, 2 x 4.5mm, 2 x 6.4mm, 1 x 8.0mm, 1 x 9.5mm diameter, all 157mm deep |
“ |
“ |
Insert Types |
Choice of three – See Accessories |
“ |
“ |
Ethernet – supporting software and USB Host |
“ |
“ |
|
CJC Accuracy |
0.35°C |
“ |
“ |
Dimensions |
384H (including handle) x 212W x 312D mm |
“ |
“ |
Power |
300 Watts |
150 Watts |
300 Watts |
Voltage |
115Vac or 230Vac 50/60Hz |
“ |
“ |
Weight |
14kg |
10.2kg |
8kg |
1 |
In ambient of 20°C: Min. Temperature is 65°C Below Ambient, Absolute Minimum -55°C |
||
2 |
In ambient of 20°C: Min. Temperature is 55°C Below Ambient, Absolute Minimum -45°C |
||
3 |
In ambient of 20°C |
||
4 |
Dry Block Mode only: Comparing 4.5mm Well to Controller Display Value |